Operation 3000ZA2G


Teledyne

Analytical

Instruments

42

4.4.1.2

M

ANUAL

M

ODE

Z

EROING

Press Zero to enter the Zero function. The screen that appears

allows you to select between automatic and manual zero calibration. Use
the ▲/▼ keys to toggle between AUTO and MAN zero settling. Stop
when MAN appears, blinking, on the display.

Zero: Settling: Man

To Begin

Press Enter to begin the zero calibration.
After the zero calibration completes its cycle, the information is

stored in the microprocessor, and the instrument automatically returns to
the Analyze mode.

4.4.2 Span Cal

The Span button on the front panel is used to span calibrate the

analyzer. Span calibration can be performed using the automatic mode,
where an internal algorithm compares consecutive readings from the
sensor to determine when the output matches the span gas concentration.
Span calibration can also be performed in manual mode, where the
operator determines when the span concentration reading is acceptable
and manually exits the function.

4.4.2.1

A

UTO

M

ODE

S

PANNING

Press Span to enter the span function. The screen that appears

allows you to select whether the span calibration is to be performed
automatically or manually. Use the ▲/▼ arrow keys to toggle between
AUTO and MAN span settling. Stop when AUTO appears, blinking, on
the display.

Span: Settling: AUTO

For Next


Press Enter to move to the next screen.