Operation 3000ZA2G


Teledyne

Analytical

Instruments

44

Span: Settling:MAN

For Next


Use the ▲/▼ keys to toggle between AUTO and MAN span

settling. Stop when MAN appears, blinking, on the display. Press Enter
to move to the next screen.

Press Enter to move to the next screen.

Calib. Holding time

Cal hold: 5 min


This menu allows the operator to set the time the analyzer should

be held in the auto span mode. It does not affect anything in Manual
Mode. Just press Enter to continue.

Span Val: 000008.00

Span Mod #


Press ▲ () to permit modification (Mod #) of span value.
Use the arrow keys to enter the oxygen concentration of the span

gas you are using (209000.00 if you are using air). The ◄►arrows
choose the digit, and the ▲/▼ arrows choose the value of the digit.

Press Enter to enter the span value into the system and begin the

span calibration.

Once the span has begun, the microprocessor samples the output at

a predetermined rate. It calculates the difference between successive
samplings and displays this difference as Slope on the screen. It takes
several seconds for the first Slope value to display. Slope indicates rate
of change of the Span reading. It is a sensitive indicator of stability.