Operation 3000ZA2G


Teledyne

Analytical

Instruments

32

will be a three-minute delay before the analog outputs and the range ID
contacts start tracking again.

The concentration alarms freeze on their last state before entering

calibration regardless of selecting HOLD or TRACK. But, when HOLD
is selected the concentration alarms will remain frozen for the time
displayed in the second line of the TRAK/HLD menu after the analyzer
returns to the Analyze mode.

The factory default is three minutes, but the delay time is program-

mable. To adjust to delay time use the ◄► arrow keys. When the time
displayed on the second line blinks, it can be adjusted by pressing the
▲/▼ keys to increase or decrease its value. The minimum delay is 1
minute, the maximum is 30.

This preference is stored in non-volatile memory so that it is

recovered if power is removed from the instrument.

4.3.2 Setting up an Auto-Cal

When proper automatic valving is connected (see Chapter 3,

Installation), the analyzer can cycle itself through a sequence of steps
that automatically zero and span the instrument.

Note: If you require highly accurate Auto-Cal timing, use external

Auto-Cal control where possible. The internal clock in the
Model 3000ZA2G is accurate to 2-3 %. Accordingly,
internally scheduled calibrations can vary 2-3 % per day.

To setup an Auto–Cal cycle:
Choose System from the Function buttons. The LCD will display

five sub functions.

TRAK/HLD Auto—Cal

PSWD Logout More

Use ◄►arrows to blink

Auto—Cal

, and press Enter. A new screen

for Span/Zero set appears.

Span OFF Nxt: 0d 0h

Zero OFF Nxt: 0d 0h