beautypg.com

Teledyne 3000ZA2G - Zirconium oxide analyzer User Manual

Page 61

background image

Trace Oxygen Analyzer

Operation

Teledyne

Analytical

Instruments

49

4.6.1 Setting the Analog Output Ranges

To set the ranges, enter the range function mode by pressing the

Range

button on the front panel.

L—10

M—1000

H—250000 Mode—AUTO


Use the ◄►arrow keys to blink the range to be set: low (L),

medium (M), or high (H).

Use the ▲/▼ arrow keys to enter the upper value of the range (all

ranges begin at 0 ppm). Repeat for each range you want to set. Press
Enter

to accept the values and return to Analyze mode. (See note

below.)

Note: The ranges must be increasing from low to high, for

example, if range 1 is set as 0–100 ppm and range 2 is set
as 0–1,000 ppm, range 3 cannot be set as 0– 500 ppm
since it is lower than range 2.

Ranges, alarms, and spans are always set in ppm units (over the entire

0-250,000 ppm range), even though all concentration-data outputs change
from ppm units to percent when the concentration is above 10,000 ppm.

4.6.2 Fixed Range Analysis

The auto ranging mode of the instrument can be overridden, forcing

the analyzer DC outputs to stay in a single predetermined range.

To switch from auto ranging to fixed range analysis, enter the range

function by pressing the Range button on the front panel.

Use the ◄►arrow keys to move the blinking over AUTO.
Use the ▲/▼ arrow keys to switch from AUTO to FX/LO, FX/MED,

or FX/H I to set the instrument on the desired fixed range (low, medium,
or high).

L—10

M—1000

H—250000

Mode—FX/ LO

or