background image

Trace Oxygen Analyzer

Operation

Teledyne

Analytical

Instruments

51

ppm O

2

Voltage Signal

Output (VDC)

Current Signal

Output (mA dc)

0 0.0

4.0

10 0.1

5.6

20 0.2

7.2

30 0.3

8.8

40 0.4 10.4
50 0.5 12.0
60 0.6 13.6
70 0.7 15.2
80 0.8 16.8
90 0.9 18.4

100 1.0 20.0

The analog output signal has a voltage which depends on the

oxygen concentration AND the currently activated analysis range. To
relate the signal output to the actual concentration, it is necessary to
know what range the instrument is currently on, especially when the
analyzer is in the auto ranging mode.

To provide an indication of the range, a second pair of analog

output terminals is used. They generate a steady preset voltage (or
current when using the current outputs) to represent a particular range.
The following table gives the range ID output for each analysis range:

Range

Voltage (V) Current (mA)

LO 0.25

8

MED 0.50

12

HI 0.75

16

CAL (0-25%)

1.00

20