Installation 3000ZA2G


Teledyne

Analytical

Instruments

26

Table 3-5: Commands via RS-232 Input

Command Description

as

Immediately starts an auto span.

az

Immediately starts an auto zero.

st

Toggling input. Stops/Starts any
status message output from the
RS-232, until st is sent
again.


The RS-232 protocol allows some flexibility in its implementation.

Table 3-6 lists certain RS-232 values that are required by the 3000ZA2G
implementation.

Table 3-6: Required RS-232 Options

Parameter Setting

Baud 2400

Byte 8

bits

Parity none

Stop

Bits 1

Message Interval

2 seconds.

3.4 Testing the System

Before applying power to the instrument:

Check the integrity and accuracy of the gas connections.
Make sure there are no leaks.

Check the integrity and accuracy of the electrical
connections. Make sure there are no exposed conductors.

Check that inlet sample pressure is within the accepted range
(see Section 3.3.1).

Power up the system, and test it by repeating the Self-
Diagnostic Test as described in Chapter 4, Section 4.3.5.