Trace Oxygen Analyzer

Appendix

Teledyne

Analytical

Instruments

83

Appendix

A-1 Specifications

Packaging: General

Purpose

• Flush panel mount (Standard)

• Relay rack mount. Contains either one

or two instruments in one 19" relay rack
mountable plate (Optional).

Sensor: Teledyne B-2CXL trace analysis Micro-

fuel Cell.

Insta-Trace Micro-fuel Cell (option)

Cell Block: 316 stainless steel.

Sample System: All wetted parts of 316 stainless steel.

90 % Response Time: 65 seconds at 25°C (77°F) on 10 ppm

range.

100 seconds on 1 ppm range.

Ranges: Three user definable ranges from 0–1 ppm

to 0–250,000 ppm, plus air calibration
range of 0-250,000 ppm (25%).

Autoranging with range ID output.

Alarms: One system-failure alarm contact to detect

power failure or sensor zero failure.

Two adjustable concentration threshold

alarm contacts with fully programmable
setpoints.

Displays: 2-line by 20-character, VFD screen, and

one 5 digit LED display.

Digital Interface: Full duplex RS-232 communications port.