Installation 3000TA-XL-EU


Teledyne

Analytical

Instruments

28

1 ID, Medium range is assigned to Range 2 ID, and High range is
assigned to Range 3 ID. The fourth range is reserved for the Air Cal
Range (25%). Table 3-4 lists the pin connections.

Table 3-4: Range ID Relay Connections

Pin Function

21

Range 1 ID Contact

38

Range 1 ID Contact

22

Range 2 ID Contact

39

Range 2 ID Contact

19

Range 3 ID Contact

18

Range 3 ID Contact

34

Range 4 ID Contact (Air Cal)

35

Range 4 ID Contact (Air Cal)

3.3.2.2.7 Network I/O

A serial digital input/output for local network protocol. At this

printing, this port is not yet functional. It is to be used for future options
to the instrument. Pins 13 (+) and 29 (–).

3.3.2.2.8 Remote Valve Connections

The 3000TA-XL-EU is a single-chassis instrument, which has no

Remote Valve Unit. Instead, the Remote Valve connections are used as
a method for directly controlling external sample/zero/span gas valves.
See Figure 3-5.