beautypg.com

9 signal output – Teledyne 3000TA-XLEU - Trace oxygen analyzer User Manual

Page 76

background image

Operation 3000TA-XL-EU


Teledyne

Analytical

Instruments

60

4.9 Signal Output

The standard Model 3000TA-XL-EU Trace Oxygen Analyzer is

equipped with two 0–1 VDC analog output terminals accessible on the
back panel (one concentration and one range ID), and two isolated 4–20
mA DC current outputs (one concentration and one range ID).

See Rear Panel in Chapter 3, Installation, for illustration.
The signal output for concentration is linear over the currently

selected analysis range. For example, if the analyzer is set on range that
was defined as 0–100 ppm O

2

, then the output would be:

Table 4-1: Output Signals

Voltage Signal Current Signal

ppm

O2

Output (VDC)

Output (mA DC)

0

0.0 4.0

10

0.1 5.6

20

0.2 7.2

30

0.3 8.8

40

0.4 10.4

50

0.5 12.0

60

0.6 13.6

70

0.7 15.2

80

0.8 16.8

90

0.9 18.4

100 1.0 20.0


The analog output signal has a voltage which depends on the

oxygen concentration AND the currently activated analysis range. To
relate the signal output to the actual concentration, it is necessary to
know what range the instrument is currently on, especially when the
analyzer is in the autoranging mode.

To provide an indication of the range, a second pair of analog

output terminals are used. They generate a steady preset voltage (or