Operation 3000TA-XL-EU


Teledyne

Analytical

Instruments

62

Ethernet web server coprocessor. To set the IP address, enter the system
menu:

Ver SysTest Sample

AL-Hld Temp-Set IP

Select IP and press Enter. The last IP address setting will be shown:

IP NUMBER

172.016.032.254

Adjust each field separated by a point using the Up/Down keys and

hop to the next field by using the Right/Left arrow keys. Once the
address is set, press the Enter key. The 3000 microcontroller will send
the new address to the Ethernet Web Enabled co-processor.

The 3000-I series analyzer microcontroller does not query the

Ethernet Web Enabled co-processor on its current IP address. In other
words, The IP address displayed is not necessarily the IP address of the
Ethernet Web Enabled co-processor, Consider this if when replacing the
Ethernet PCB.

4.11 Cold Boot to Return to Default Settings

Occasionally, it may be necessary to return the instrument to its

default settings. This can be done using a cold boot as follows:

Turn the power off to the instrument, wait 5 seconds, and then turn

the power back on.

On the initial screen you will be prompted to

Press System for

Defaults

. Doing so will revert the analyzer back to the default parameters

as set at the factory. The default settings are:

Ranges:

LO = 1 ppm, MED = 10 ppm, HI = 100 ppm.

Auto Ranging: ON

Alarm Relays: Defeated, 10.00 ppm, HI, Not failsafe, Not

latching.

Zero:

Auto, every 0 days at 0 hours.

Span: Auto, at 000,008.00 ppm, every 0 days at 0

hours.