3.4 Installing the Micro-fuel Cell

The Model 3000TA-XL-EU can be equipped with either the

standard B-2CXL Micro-fuel Cell or as an option, Teledyne’s Insta-
Trace cell. An existing instrument can also be retrofitted to use the Insta-
Trace cell using Teledyne retrofit kit P/N C-71792-A. This kit contains a
cell holder and an Insta-Trace cell.

3.4.1 B-2CXL Micro-fuel Cell

The B-2CXL Micro-fuel Cell is not installed in the cell block when

the instrument is shipped. It must be installed before the analyzer is
placed in service.

The Micro-Fuel cell is located inside the stainless steel cell block

behind the front panel (see Figure 3-7).

To install the cell:

1. Remove power to the instrument by unplugging the power

cord at the power source.

2. Open the front panel door by pressing the release button on

the top right corner of the door all the way in with a narrow
gauge tool.

3. With one hand placed underneath the cell block ready to

catch the micro-fuel cell , lift up on the stainless steel gate in
front of the cell block. This releases the cell and cell holder
from the block. The cell and holder will fall out in your hand.

CAUTION:

THE CELL IS SHIPPED SEPARATELY INSIDE TWO
BAGS FILLED WITH OXYGEN-FREE INERT GAS.
EXTREME CARE MUST BE TAKEN TO ENSURE
THAT THE CELL IS EXPOSED TO AIR FOR A VERY