Operation 3000TA-XL-EU
Teledyne
Analytical
Instruments
40
4.3 The System Function
The sub functions of the System function are described below.
Specific procedures for their use follow the descriptions:
Auto-Cal: Used to define an automatic calibration sequence
and/or start an Auto-Cal.
PSWD: Security can be established by choosing a 5 digit password
(PSWD) from the standard ASCII character set. (See Installing or
Changing the Password, below, for a table of ASCII characters
available.) Once a unique password is assigned and activated, the
operator MUST enter the UNIQUE password to gain access to set-up
functions which alter the instrument's operation, such as setting the
instrument span or zero setting, adjusting the alarm setpoints, or
defining analysis ranges.
After a password is assigned, the operator must log out to
activate it. Until then, anyone can continue to operate the
instrument without entering the new password.
Only one password can be defined. Before a unique
password is assigned, the system assigns TETAI by
default. This allows access to anyone. After a unique
password is assigned, to defeat the security, the password
must be changed back to TETAI.
Logout: Logging out prevents unauthorized tampering with
analyzer settings.
More: Select and enter More to get a new screen with additional
sub functions listed.
Self–Test: The instrument performs a self-diagnostic test to check
the integrity of the power supply, output boards and amplifiers.
Version: Displays Manufacturer, Model, and Software Version of
the instrument.
Show Negative: The operator selects whether the display can
show negative oxygen readings or not.
TRAK/HLD: The operator sets whether the instrument analog
outputs track the concentration change during calibration and
sets a time delay for the concentration alarms after calibration.