beautypg.com

Teledyne 3000TA-XLEU - Trace oxygen analyzer User Manual

Page 42

background image

Installation 3000TA-XL-EU


Teledyne

Analytical

Instruments

26

43

Threshold Alarm 2, normally open contact

36

System Alarm, normally closed contact

20

System Alarm, moving contact

37

System Alarm, normally open contact

3.3.2.2.3 Digital Remote Cal Inputs

These inputs accept 0 V (off) or 24

VDC (on) inputs for remote

control of calibration. (See Remote Calibration Protocol below.) See
Table 3-3 for pin connections.
Zero: Floating input. 5 to 24 V input across the + and – pins puts the

analyzer into the Zero mode. Either side may be grounded at
the source of the signal. 0 to 1 volt across the terminals allows
Zero

mode to terminate when done. A synchronous signal must

open and close the external zero valve appropriately. See
Remote Probe Connector. (The –C option internal valves
operate automatically.)


Span: Floating input. 5 to 24 V input across the + and – pins puts the

analyzer into the Span mode. Either side may be grounded at
the source of the signal. 0 to 1 volt across the terminals allows
Span

mode to terminate when done. A synchronous signal

must open and close external span valve appropriately. See
Figure 3-5 Remote Probe Connector. (The –C option internal
valves operate automatically.)

3.3.2.2.4 Cal Contact

This relay contact is closed while analyzer is spanning and/or

zeroing. (See Remote Calibration Protocol below.)

Table 3-3: Remote Calibration Connections

Pin Function

9

+ Remote Zero

11

– Remote Zero

10

+ Remote Span

12

– Remote Span