Installation 3000TA-XL-EU
Teledyne
Analytical
Instruments
24
0–1 V DC % of Range: Voltage rises linearly with increasing
oxygen, from 0 V at 0 ppm to 1 V at full
scale ppm. (Full scale = 100% of
programmable range.)
0–1 VDC Range ID:
0.25 V = Low Range
0.5 V = Medium Range
0.75 V = High Range
1 V = Air Cal Range.
4–20 mA DC % Range: Current increases linearly with increasing
oxygen, from 4 mA at 0 ppm to 20 mA at
full scale ppm. (Full scale = 100% of
programmable range)
4–20 mA DC Range ID: 8 mA = Low Range
12 mA = Medium Range
16 mA = High Range
20 mA = Air Cal Range.
Table 3-1: Analog Output Connections
Pin
Function
3
+ Range ID, 4-20 mA, floating
4
– Range ID, 4-20 mA, floating
5
+ % Range, 4-20 mA, floating
6
– % Range, 4-20 mA, floating
8
+ Range ID, 0-1 VDC
23
– Range ID, 0-1 VDC, negative ground
24
+ % Range, 0-1 VDC
7
– % Range, 0-1 VDC, negative ground
3.3.2.2.2 Alarm Relays
The nine alarm-circuit connector pins connect to the internal alarm
relay contacts. Each set of three pins provides one set of Form C relay
contacts. Each relay has both normally open and normally closed contact
connections. The contact connections are shown in Table 3-2. They are
capable of switching up to 3 amperes at 250 V ac into a resistive load.
The connectors are: