background image

Trace Oxygen Analyzer

Operation

Teledyne

Analytical

Instruments

37

Operation

4.1 Introduction

Once the analyzer has been installed, it can be configured for your

application. To do this you will:

Set system parameters.

 Establish a security password, if desired, requiring

Operator to log in.

 Establish and start an automatic calibration cycle, if

desired.

Calibrate the instrument.

Define the three user selectable analysis ranges, then choose
autoranging or select a fixed range of analysis, as required.

Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc).

Before you configure your 3000TA-XL-EU, these default values

are in effect:

Ranges: LO = 1 ppm, MED = 10 ppm, HI = 100 ppm.

Auto Ranging: ON

Alarm Relays: Defeated, 10.00 ppm, HI, Not failsafe, Not

latching.

Zero: Auto, every 0 days at 0 hours.

Span: Auto, at 000,008.00 ppm, every 0 days at 0

hours.

If you choose not to use password protection, the default password

is automatically displayed on the password screen when you start up,
and you simply press Enter for access to all functions of the analyzer.