Trace Oxygen Analyzer

Installation

Teledyne

Analytical

Instruments

19

Installation

Installation of the Model 3000TA-XL-EU Analyzer includes:

1. Unpacking
2. Mounting
3. Gas connections
4. Electrical connections
5. Installing the Micro-Fuel Cell
6. Testing the system.

3.1 Unpacking the Analyzer

The analyzer is shipped with all the materials you need to install

and prepare the system for operation. Carefully unpack the analyzer and
inspect it for damage. Immediately report any damage to the shipping
agent.

3.2 Mounting the Analyzer

The Model 3000TA-XL-EU is for indoor use in a general purpose

area. It is NOT for hazardous environments of any type.

The standard model is designed for flush panel mounting. Figure 3-1

is an illustration of the 3000TA-XL-EU standard front panel and
mounting bezel. There are four mounting holes—one in each corner of the
rigid frame. The drawings at the back of this manual includes a panel
cutout diagram.

On special order, a 19" rack-mounting panel can be provided. For

rack mounting, one or two 3000XL series analyzers can be flush-panel
mounted on the rack panel. See Appendix for dimensions of the
mounting panel.