Trace Oxygen Analyzer

Operation

Teledyne

Analytical

Instruments

47

4.3.5 System Self-Diagnostic Test

The Model 3000TA-XL-EU has a built-in self-diagnostic testing

routine. Pre-programmed signals are sent through the power supply,
output board and sensor circuit. The return signal is analyzed, and at the
end of the test the status of each function is displayed on the screen,
either as OK or as a number between 1 and 3. (See System Self
Diagnostic Test
in Chapter 6 for number code).

The self diagnostics are run automatically by the analyzer

whenever the instrument is turned on, but the test can also be run by the
operator at any time. In addition, web enabled instruments can initiate a
self diagnostic test remotely from a computer. See Section 5.6.

To initiate a self diagnostic test during operation:
Press the System button to start the System function.

TRAK/HLD Auto—Cal
PSWD Logout More

Use the ◄► arrow keys to blink

More

, then press Enter.

Version Self—Test

Use the ◄► arrow keys again to move the blinking to the

Self–Test

function. The screen will follow the running of the diagnostic.

RUNNING DIAGNOSTIC
Testing Preamp — 83

During preamp testing there is a countdown in the lower right

corner of the screen. When the testing is complete, the results are
displayed.

Power: OK Analog: OK
Preamp: 3

The module is functioning properly if it is followed by OK. A

number indicates a problem in a specific area of the instrument. Refer to
Section 6.7 Troubleshooting for number-code information. The results
screen alternates for a time with:

Press Any Key
To Continue...

Then the analyzer returns to the initial System screen.