beautypg.com

Teledyne 3000TA-XLEU - Trace oxygen analyzer User Manual

Page 65

background image

Trace Oxygen Analyzer

Operation

Teledyne

Analytical

Instruments

49

4.4.1 Zero Cal

The B-2CXL cell has a zero offset of less than 0.2 ppm oxygen.

Normally, the offset slowly decreases during the first 7 to 10 days of
operation, and is expected to reach a steady value after this time.

Generally, the value of the zero offset is part of the oxygen reading

of the sample gas as shown by the analyzer readout. As an example, a
reading of 0.5 ppm oxygen may include 0.4 ppm oxygen in the sample
gas and a 0.1 ppm zero offset.

The determination of the zero offset requires the use of oxygen free

gas to the analyzer. We recommend the use of nitrogen gas with a
scrubber to assure oxygen levels below 0.02 ppm.

The user may decide to eliminate the zero offset for improved

accuracy. If the low range accuracy needs to be better than 100 ppb, a
zero calibration should be performed. Thus, the analyzer is equipped
with a zero calibration function. However, we do not recommend
carrying out the cal zero during the first 10 days of the operation of the
cell.

As long as the unit has not been contaminated by a liquid, the

default zero of this instrument will be stable (± 100 ppb or less) in
excess of 10 years.

The Zero button on the front panel is used to enter the zero

calibration function. Zero calibration can be performed in either the
automatic or manual mode. In the automatic mode, an internal algorithm
compares consecutive readings from the sensor to determine when the
output is within the acceptable range for zero. In the manual mode, the
operator determines when the reading is within the acceptable range for
zero. Make sure the zero gas is connected to the instrument. If you get a

CELL FAILURE

message skip to Section 4.4.1.3.

4.4.1.1

A

UTO

M

ODE

Z

EROING

Press Zero to enter the zero function mode. The screen allows you

to select whether the zero calibration is to be performed automatically or
manually. Use the ▲▼

arrow keys to toggle between

AUTO

and

MAN

zero settling. Stop when

AUTO

appears, blinking, on the display.

Zero: Settling: AUTO
To Begin

Press Enter to begin zeroing.