Maintenance – Teledyne 3000TA-XLEU - Trace oxygen analyzer User Manual

Page 89

background image

Trace Oxygen Analyzer

Maintenance

Teledyne

Analytical

Instruments

73

Maintenance

6.1 Routine Maintenance

Aside from normal cleaning and checking for leaks at the gas

connections, routine maintenance is limited to replacing Micro-Fuel
cells and fuses, and recalibration. For recalibration, see Section 4.4
Calibration.

WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS

MANUAL.


6.2 Cell Replacement

The B-2CXL Micro-Fuel Cell and Insta-Trace Micro-fuel Cells are

sealed electrochemical transducers with no electrolyte to change or
electrodes to clean. When the cell reaches the end of its useful life, it is
replaced. The spent fuel cell should be discarded according to local
regulations. This section describes fuel cell care as well as when and
how to replace it. Except where specifically noted, the information
pertains to both classes of cells.

6.2.1 Storing and Handling Replacement Cells

To have a replacement cell available when it is needed, TAI

recommends that one spare cell be purchased 9-10 months after
commissioning the 3000TA-XL-EU, or shortly before the end of the
cell's one year warranty period.

CAUTION:

DO NOT STOCKPILE CELLS. THE WARRANTY
PERIOD STARTS ON THE DAY OF SHIPMENT.