background image

Trace Oxygen Analyzer

Web Implementation

Teledyne

Analytical

Instruments

67

5.2.2 Remote Zero

To perform a remote zero, click the Zero button at the top of the

HTML page. A zero calibration window will open. In this window the
operator can set zero operation to be automatic or manual.

Once set, click the Start button. If an auto calibration was chosen,

the operator must wait for the analyzer to finish the calibration. In the
main data window, the field 'mode' will switch from 'analyzing' to
'getting zero'. Check the message window to find out when calibration
is finished. If manual calibration was chosen, the operator must click the
Finish

button to force the instrument calibration.

Figure 5-3: Zero Calibration Window