Trace Oxygen Analyzer

Maintenance

Teledyne

Analytical

Instruments

81

The 3000TA-XL contains the following major components:

Analysis Section

Micro Fuel Cell (B-2CXL)

Stainless steel cell block

Sample system

Power Supply

Microprocessor

Displays

5 digit LED meter

2 line, 20 character, alphanumeric, VFD display

RS-232 Communications Port

See the drawings in the Drawings section in back of this manual for

details.

6.6 Cleaning

If the instrument is unmounted at time of cleaning, disconnect the

instrument from the power source. Close and latch the front-panel access
door. Clean outside surfaces with a soft cloth dampened slightly with
plain clean water. Do not use any harsh solvents such as paint thinner or
benzene.

For panel-mounted instruments, clean the front panel as prescribed

in the above paragraph. DO NOT wipe front panel while the instrument
is controlling your process.

6.7 Troubleshooting

Problem:

Erratic readings of the Oxygen concentration as reported by

the analyzer.

Possible Cause:

The analyzer may have been calibrated in an inaccurate fashion.