damage may occur. For cell pressures above atmospheric, caution must
be exercised to avoid over pressuring the cell holder.

FLOW CONTROL VALVE:

All 3000XL series analyzers have a flow control device. This

device is a needle valve, and serves two functions in the sample path:

1. It limits the flow rate of the sample through the analyzer. The

needle valve is chosen to operate over a specific range of
pressures and provide a useable flow rate over that range.

2. The needle valve provides a pressure drop. This device is

selected to provide the only significant pressure drop in the
sample path.

BYPASS:

To improve the system response, a bypass can be added to increase

the sample flow rate to the analyzer by a factor of ten. A by-pass
provides a sample flow path around the analyzer of 2 - 18 SCFH
typically.

CONVERSIONS:

1 PSI =

2.04 INCHES OF MERCURY (in. Hg.)

1 SCFH =

0.476 SLPM