5 Maintenance

Model 2020

5-2

Teledyne Analytical Instruments

2

Amplifier output doesn't match test input

3

Both Failed

>3

Open gain resistor

Cell

0

OK

1

Failed (open filament, short to ground, no
power.)

2

Unbalance (deterioration of filaments,
blocked tube)

5.3

Fuse Replacement

The 2020 requires two 5 x 20 mm, 4 A, T type (Slow Blow) fuses.

The fuses are located inside the explosion proof housing on the Electrical
Connector Panel, as shown in Figure 5-1. To replace a fuse:

1. Disconnect the Unit from its power source.

2. Place a small screwdriver in the notch in the fuse holder cap,

push in, and rotate 1/4 turn. The cap will pop out a few
millimeters. Pull out the fuse holder cap and fuse, as shown in
Figure 5-1.

Figure 5-1: Removing Fuse Cap and Fuse from Holder

3. Replace fuse by reversing process in step 1.

Remove Power to the instrument before changing fuses.