beautypg.com

Operational theory – Teledyne 2020 - Explosion proof thermal conductivity analyzer User Manual

Page 13

background image

2-1

Thermal Conductivity Analyzer

Operational Theory 2

Teledyne Analytical Instruments

Operational Theory

2.1

Introduction

The analyzer is composed of two subsystems:

1. Thermal Conductivity Sensor
2. Electronic Signal Processing, Display and Control.

The sensor is a thermal conductivity comparator that continuously

compares the thermal conductivity of the sample gas with that of a refer-
ence gas having a known conductivity.

The electronic signal processing, display and control subsystem

simplifies operation of the analyzer and accurately processes the sampled
data. A microprocessor controls all signal processing, input/output, and
display functions for the analyzer.

2.2

Sensor Theory

For greater clarity, Figure 2-1 presents two different illustrations, (a)

and (b), of the operating principle of the thermal conductivity cell.

2.2.1 Sensor Configuration

The thermal conductivity sensor contains two chambers, one for the

reference gas of known conductivity and one for the sample gas. Each
chamber contains a pair of heated filaments. Depending on its thermal
conductivity, each of the gases conducts a quantity of heat away from the
filaments in its chamber. See Figure 2-1(a).

The resistance of the filaments depends on their temperature. These

filaments are parts of the two legs of a Wheatstone bridge circuit that
unbalances if the resistances of its two legs do not match. See Figure
2-1(b).