Thermal Conductivity Analyzer

Operation 4

4-11

Teledyne Analytical Instruments

Then the analyzer returns to the initial System screen.

4.3.6 The Model Screen

Scroll through the MAIN MENU to MODEL and Enter. The screen

displays the manufacturer, model, and software version information.

4.3.7 Checking Linearity with ALGORITHM

Use UP/DOWN control to select

ALGORITHM, and

Enter

.

sel rng to set algo:

—> Ш1 Ш2 Ш3 <—

Use the UP/DOWN Control to select the range: 01, 02, or 03. Then

press

Enter

.

Gas Use: H2 — N2

Range: Ø — 10%

Enter again.

Algorithm setup:

VERIFY SET UP

Use UP/DOWN key to select and

Enter

VERIFY to check whether the

linearization has been accomplished satisfactorily.

Dpt INPUT OUTPUT

Ш Ш.ШШ Ш.ШШ

The leftmost digit (under

Dpt) is the number of the data point being

monitored. Use the UP/DOWN key to select the successive points.

The

INPUT value is the input to the linearizer. It is the simulated output

of the analyzer. You do not need to actually flow gas.

The

OUTPUT value is the output of the linearizer. It should be the

ACTUAL concentration of the span gas being simulated.

If the

OUTPUT value shown is not correct, the linearization must be

corrected.

ESCAPE

to return to the previous screen. Select and Enter

SET

UP to Calibration Mode screen.

Select algorithm

mode : AUTO

There are two ways to linearize:

AUTO and MANUAL: The auto mode

requires as many calibration gases as there will be correction points along