3 Installation

Model 2020

3-4

3.3.2 Fuse Installation

The fuse block, at the right of the power cord receptacle, accepts US

or European size fuses. A jumper replaces the fuse in whichever fuse
receptacle is not used. Be sure to install the proper fuse as part of installa-
tion. (See Fuse Replacement in chapter 5, maintenance.)

3.3.3 Voltage Selections

There is a switch on the interface board, inside the instrument, that

selects the working voltage between 230/115 VAC.

Voltage Selector Switch

Make sure the switch is in the proper position before
powering the instrument
.

3.3.4 Analog Outputs

There are four DC output signal connectors on the panel. There are

two wires per output with the polarity noted. See Figure 3-4. The outputs
are:

0–1 V dc % of Range: Voltage rises linearly with increasing concentra-

tion, from 0 V at 0 concentration to 1 V at full
scale. (Full scale = 100% of programmable range.)

0–1 V dc Range ID:

0.25 V = Range 1, 0.5 V = Range 2, 0.75 V =
Range 3, 1 V = Cal Range.

4–20 mA dc % Range: Current rises linearly with concentration, from 4

mA at 0 concentration to 20 mA at full scale. (Full
scale = 100% of programmable range.)

4–20 mA dc Range ID: 8 mA = Range 1, 12 mA = Range 2, 16 mA =

Range 3, 20 mA = Range 4.