4 Operation

Model 2020

4-18

Teledyne Analytical Instruments

In failsafe mode, the alarm relay de-energizes in an alarm
condition. For non-failsafe operation, the relay is energized in
an alarm condition. You can set either or both of the
concentration alarms to operate in failsafe or non-failsafe mode.

3. Are either of the alarms to be latching?

In latching mode, once the alarm or alarms trigger, they will
remain in the alarm mode even if process conditions revert back
to non-alarm conditions. This mode requires an alarm to be
recognized before it can be reset. In the non-latching mode, the
alarm status will terminate when process conditions revert to
non-alarm conditions.

4. Are either of the alarms to be defeated?

The defeat alarm mode is incorporated into the alarm circuit so
that maintenance can be performed under conditions which
would normally activate the alarms.

The defeat function can also be used to reset a latched alarm.
(See procedures, below.)

If you are using password protection, you will need to enter your

password to access the alarm functions. Follow the instructions in section
4.3.3 to enter your password. Once you have clearance to proceed, enter
the

Alarm

function.

Note: If all ranges are for the same application, set any one of them

will automatically set the others.

Press the

Alarm

button on the front panel to enter the

Alarm

function.

Make sure that

01 is blinking.

Sel rng to set alm:

—> Ш1 Ш2 Ш3 <—

Set up the Range 1 alarm by moving the blinking over to

01 using the

UP/DOWN arrow keys. Then

Enter

. Check the gas application and range

limits as displayed on the screen.

Gas use: H2 — N2

Range: 0 — 10 %

Press enter again to set the alarm setpoints.

Sel %/ppm alm to set

AL1—PPM AL2—PPM

Use the UP/DOWN keys to choose between % or ppm units. Then

Enter

to move to the next screen.