3 Installation

Model 2020

3-10

Table 3-4: Commands via RS-232 Input

Command

Description

as

Immediately starts an autospan.

az

Immediately starts an autozero.

rp

Allows reprogramming of the

APPLICATION (gas use) and

ALGORITHM (linearization)

System

functions.

st

Toggling input. Stops/Starts any status message output
from the RS-232, until st is sent again.

rm1

Range manual 1

rm2

Range manual 2

rm3

Range manual 3

rm4

Range manual CAL

ra

Range auto

Implementation: The RS-232 protocol allows some flexibility in its

implementation. Table 3-4 lists certain RS-232 values that are required by
the Model 2020 implementation.

Table 3-5: Required RS-232 Options

Parameter

Setting

Baud

2400

Byte

8 bits

Parity

none

Stop Bits

1

Message Interval

2 seconds

3.3.10 Remote Probe Connector

The 2020 is a single-chassis instrument, which has no Remote Probe

Unit. Instead, the Remote Probe connector is used as another method for
controlling external sample/zero/span gas valves. See Figure 3-6.