beautypg.com

Teledyne 2020 - Explosion proof thermal conductivity analyzer User Manual

Page 49

background image

Thermal Conductivity Analyzer

Operation 4

4-15

Teledyne Analytical Instruments

4.4.1.3 Cell Failure

Cell failure in the 2020 is usually associated with inability to zero the

instrument with a reasonable voltage differential across the Wheatstone
bridge. If this should ever happen, the 2020 system alarm trips, and the
VFD displays a failure message.

Cell cannot be balanced

Check your zero gas

Before replacing the sensor:

a. Check your zero gas to make sure it is within specifications.

b. Check for leaks downstream from the sensor, where contamina-

tion may be leaking into the system.

c. Check flowmeter to ensure that the flow is no more than

200SCCM

d. Check temperature controller board.

e. Check gas temperature.

If none of the above as indicated, the sensor may need to be replaced.

Check warranty, and contact Analytical Instruments Customer Service.

4.4.2 Span Cal

The

Span

button on the front panel is used to span calibrate the

analyzer. Span calibration can be performed in either the automatic or
manual mode.

CAUTION: If you are spanning the Cal Range by itself (multiple

application analyzers only), use

manual mode

zeroing.

If you want to calibrate ALL of the ranges at once
(multiple application analyzers only), use

auto mode

spanning in the Cal Range.

Make sure the span gas is flowing to the instrument.

4.4.2.1

Auto Mode Spanning

Observe all precautions in sections 4.4 and 4.4.2, above. Scroll SPAN

and enter the span function. The screen that appears allows you to select
whether the span calibration is to be performed automatically or manually.