beautypg.com

Teledyne 2020 - Explosion proof thermal conductivity analyzer User Manual

Page 25

background image

Teledyne Analytical Instruments

Thermal Conductivity Analyzer

Installation 3

3-7

Actuates when self test fails.

To reset a system alarm, call out the set up menue

by scroll keys. Use UP/DOWN key to select
STANDBY function. Turn off analyzer by pressing
ENTER key. Turn analyzer back on by selecting
any key. Set ESC key twice.

Further detail can be found in chapter 4, section 4-5.

DANGEROUS VOLTAGES MAY STILL BE
PRESENT AT THIS TERMINALS EVEN IF POWER
TO THE INSTRUMENT IS REMOVED.

Figure 3-5: Types of Relay Contacts

3.3.6 Digital Remote Cal Inputs

Accept 0 V (off) or 24 V dc (on) inputs for remote control of calibra-

tion. (See Remote Calibration Protocol below.)

Zero:

Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the

Zero

mode. Either side may be

grounded at the source of the signal. A synchronous signal
must open and close the external gas control valves appro-
priately. See 3.3.9 Remote Probe Connector. (With the –C
option, the internal valves operate automatically.)

Span:

Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the

Span

mode. Either side may be

grounded at the source of the signal. A synchronous signal
must open and close the external gas control valves appro-