Installation 3

3-3

Figure 3-2: Required Front Door Clearance

3.3

Electrical Connections

Figure 3-3 shows the Model 2020 Electrical Connector Panel. There

are terminal blocks for connecting power, communications, and both
digital and analog concentration outputs.

For safe connections, ensure that no uninsulated wire extends outside

of the connectors they are attached to. Stripped wire ends must insert
completely into terminal blocks. No uninsulated wiring should be able to
come in contact with fingers, tools or clothing during normal operation.

3.3.1 Primary Input Power

The power cord receptacle and fuse block are located in the same

assembly. Insert the female plug end of the power cord into the power cord
receptacle.

DANGER:

POWER IS APPLIED TO THE INSTRUMENT'S CIR-
CUITRY AS LONG AS THE INSTRUMENT IS CON-
NECTED TO THE POWER SOURCE. THE STANDBY
FUNCTION IS FOR SWITCHING POWER ON / OFF
TO THE DISPLAY AND OUTPUTS ONLY.

The standard power supply requires a 115 V ac, 50-60 Hz power

source. If you have the -N option, you will require 220 V ac, 50-60 Hz
power.