background image

4 Operation

Model 2020

4-20

Teledyne Analytical Instruments

In the

AUTO screen, you are further allowed to select which gas appli-

cation (PREVIOUSLY defined in

APPLICATION

function) to run.

4.6.1 Manual (Select/Define Range) Screen

The Manual range-switching mode allows you to select a single, fixed

analysis range. It then allows you to redefine the upper and lower limits, for
the range.

Use UP/DOWN key to start the RANGE function, and ENTER

Select range

mode: MANUAL

Note: If all three ranges are currently defined for different applica-

tion gases, then the above screen does not display (because
mode must be manual). Instead, the VFD goes directly to the
following screen.

If above screen displays, use the UP/DOWN arrow keys to Select

MANUAL, and press

Enter.

Select range to run

—> Ш1 Ш2 Ш3 CAL<—

Use the UP/DOWN keys to select the range: 01, 02, 03, or CAL.

Then press

Enter

.

Gas use: H2 — N2

Range: Ø — 10 %

Use the ENTER key to move the range to low-end field. Use ENTER

key to move the range to high-end field. Use the UP/DOWN keys to
change the values of the fields.

Press

Escape

to return to the previous screen to select or define

another range.

Press

Enter

to return the to the

Analyze

function.

4.6.2 Auto (Single Application) Screen

Autoranging will automatically set to the application that has at least

two range setup with the same gases.

In the autoranging mode, the microprocessor automatically responds

to concentration changes by switching ranges for optimum readout sensitivi-
ty. If the upper limit of the operating range is reached, the instrument