Thermal Conductivity Analyzer

Operation 4

4-31

Teledyne Analytical Instruments

For nonlinear application, signal that produces from the termal conduc-

tivity, will not correspond to the actual gas concentration. The amplifica-
tion of each range will not agree, therefore, the gain must be preset in order
for the signal to read linearly. To set the gain, the following must be per-
formed in sequence.

NOTE: Before setting up this feature, you must have a span gas

containing 90%-100% of the lowest range of the analyzer.

1. Set unit range to lowest range.

NOTE: For output reversal, the lowest range should be range 3, else

the lowest range is range 1.

2. Connect span gas to span inlet.

3. Use UP/DOWN key to scroll to the CAL-INDEPD function.

4. Hold Escape/Enter control to ENTER position for approximately 10

seconds until the upper right connect display “ok” message.

5. Use UP/DOWN scroll to select SPAN function, then setup the

setting to span level. Press ENTER key to span.

NOTE: You must do step #5 before the analyzer return to analyze

mode. If the analyzer returns to analyze function, you must
repeat step 3-5 again.

6. Set range to nex range’

7. Repeat steps 3-5 until you reach the last range of the row.