beautypg.com

Teledyne 2020 - Explosion proof thermal conductivity analyzer User Manual

Page 59

background image

Thermal Conductivity Analyzer

Operation 4

4-25

Teledyne Analytical Instruments

Use the UP/DOWN key to increment/decrement the range number to

01, 02, 03, or CAL, and

Enter

.

Imp: H2 Bck: N2

FR:Ø TO:1Ø %

Use the UP/DOWN key to increment the respective parameters as

desired.

Use the ENTER to move from Imp: (impurity) to Bck: (background),

FR: (from—lower end of range), TO: (to—upper end of range), and PPM
or %.

Last

Enter

will accept the values and the screen will display

OFFST/INVRT:

Standard:

ESC if your application is standard. If your application requires

OFFSET or Reversal output, ENTER will set your output as OFFSET or
REVERSAL. (See special function setup section 4.9 for more information).

. (See note below.) Repeat for each range you want to set.

Note: The ranges must be increasing from low to high, for example,

if Range 1 is set to 0–10 % and Range 2 is set to 0–100 %,
then Range 3 cannot be set to 0–50 % since that makes
Range 3 lower than Range 2.

Ranges, alarms, and spans are always set in either percent or ppm

units, as selected by the operator, even though all concentration-data out-
puts change from ppm to percent when the concentration is above
9999 ppm.

Note: When performing analysis on a fixed range, if the concentra-

tion rises above the upper limit as established by the operator
for that particular range, the output saturates at 1 V dc (or 20
mA). However, the digital readout and the RS-232 output
continue to read regardless of the analog output range.

To end the session, send:

st
st

to the analyzer from the computer.