beautypg.com

Teledyne 2020 - Explosion proof thermal conductivity analyzer User Manual

Page 60

background image

4 Operation

Model 2020

4-26

Teledyne Analytical Instruments

4.8.2 The Curve Algorithm Screen

The Curve Algorithm is a linearization method. It provides from 1 to 9

intermediate points between the ZERO and SPAN values, which can be
normalized during calibration, to ensure a straight-line input/output transfer
function through the analyzer. (Before setting the alogorithm curve, each
range must be Zeroed and Spanned
).

Each range is linearized individually, as necessary, since each range

will usually have a totally different linearization requirement.

To linearize the ranges, you must first perform the four steps indicated

at the beginning of section 4.8 Programming. You will then be in the MAIN
MENU and select ALOGORITHM.

4.8.2.1 Checking the linearization

From the

MAIN MENU

screen, select

ALGORITHM, and

Enter

.

Sel rng set algo

—> Ш1 Ш2 Ш3 <—

Use the UP/DOWN key to select the range: 01, 02, or 03. Then press

Enter

.

Gas use: H2 — N2

Range: Ø -10 %
Enter again.

Algorithm setup:

VERIFY SETUP

UP/DOWN to select and

Enter

VERIFY to check whether the lineariza-

tion has been accomplished satisfactorily.

Dpt INPUT OUTPUT

Ш Ш.ШШ Ш.ШШ

The leftmost digit (under

Dpt) is the number of the data point being

monitored. Use the UP/DOWN keys to select the successive points.

The

INPUT value is the input to the linearizer. It is the simulated output

of the analyzer. You do not need to actually flow gas.

The

OUTPUT value is the output of the linearizer. It should be the

ACTUAL concentration of the span gas being simulated.

If the

OUTPUT value shown is not correct, the linearization must be

corrected.

ESCAPE

to return to the previous screen. Select and Enter

SET

UP to Calibration Mode screen.