background image

4 Operation

Model 2020

4-16

Teledyne Analytical Instruments

Use the UP/DOWN key to toggle between

AUTO and MAN span settling.

Stop when

AUTO appears, blinking, on the display.

Select span

mode: AUTO

Enter

to move to the next screen.

Span Val: 2Ш.ШШ %

To begin span

Use UP/DOWN key to change the span setting value.

ENTER will move the blinking field to units (%/ppm). Use UP/DOWN

key to select the units, as necessary. When you have set the concentration of
the span gas you are using,

Enter

to begin the Span calibration.

####.##% H2 — N2

Slope=#.### Span

The beginning span value is shown in the upper left corner of the

display. As the span reading settles, the screen displays and updates infor-
mation on

Slope. Spanning automatically ends when the span output corre-

sponds, within tolerance, to the value of the span gas concentration. Then
the instrument automatically returns to the analyze mode.

4.4.2.2

Manual Mode Spanning

Scroll Span by using UP/DOWN key and Enter to start the Span

function. The screen that appears allows you to select whether the span
calibration is to be performed automatically or manually.

Select span

mode: MANUAL

Use the UP/DOWN key to toggle between

AUTO and MAN span

setting. Stop when

MAN appears, blinking, on the display. ENTER to move

to next subfunction screen

Span Val: 2Ш.ШШ %

To begin span

Using the UP/DOWN key changes the span value, as necessary.

Enter

to move to the units field (%/ppm). Use UP/DOWN key to select unit.

Press

Enter

to enter the span value into the system and begin the span

calibration.

Once the span has begun, the microprocessor samples the output at a

predetermined rate. It calculates the difference between successive sam-