Thermal Conductivity Analyzer

Maintenance 5

5-1

Teledyne Analytical Instruments

Maintenance

5.1

Routine Maintenance

Aside from normal cleaning and checking for leaks at the gas connec-

tions, routine maintenance is limited to replacing fuses, and recalibration.
For recalibration, see Section 4.4 Calibration.

WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS

MANUAL.

5.2

System Self Diagnostic Test

Use UP/DOWN key to enter the SELF-TESY function.

The following failure codes apply:

Table 5-1: Self Test Failure Codes

Power

0

OK

1

5 V Failure

2

15 V Failure

3

Both Failed

Analog

0

OK

1

DAC A (0–1 V Concentration)

2

DAC B (0–1 V Range ID)

3

Both Failed

Preamp

0

OK

1

Zero too high