beautypg.com

Teledyne 2020 - Explosion proof thermal conductivity analyzer User Manual

Page 30

background image

Teledyne Analytical Instruments

3 Installation

Model 2020

3-12

Figure 3-8: Gas Connections to the Basic Unit

There are no gas control valves inside the main chassis. A sample

system must be provided for introduction of zero and span gas, as well as
sample gas, into the sample path, and for controlling the flowrates through
the sample and reference paths of the analyzer.

If you have purchased a gas selector panel from Analytical Instru-

ments, the drawings at the back of this manual will contain a dimension
drawing, with the modified cutout and hole pattern for mounting, and a
drawing and/or addendum showing the gas connections. Figure 3-9 is an
example showing a manual-valve panel with three valves (for sample, span
and zero gases) and two flowmeters (one for reference and one for sample,
span and zero gases).