beautypg.com

Teledyne 2010A - Split architecture thermal conductivity analyzer User Manual

Page 90

background image

2 Installation

Model 2010A

2-4: Part II

Teledyne Analytical Instruments

vented to the same area, both VENT lines must vent to areas with equal
ambient pressures, and pressures must vary no more than the normal baro-
metric changes.

2-6

SAMPLE Gas

In the standard model, sample and calibration gases are introduced

through the

SAMPLE

fitting. The gases must be Tee'd into the Sample inlet

with appropriate valves.

The gas pressure in should be well regulated. The sample line pressure

regulator should be installed as close to the sample line as possible to mini-
mize sample line lag time.

If greater flow is required for improved response time, install a bypass

in the sampling system upstream of the analyzer input.

2-7

REFERENCE Gas

A gas of fixed composition is needed as a reference to which the

sample gas will be compared. The reference gas is normally selected to
represent the main background gas of the analysis.

For most applications, a constant supply of reference gas flowing at the

same rate as the sample is required for best results. However, in many cases
the flow of reference gas can be slowed to about 0.08 scfh (40 cc/min) with
good results.

For some applications, an optional sealed air reference is installed. In

sealed-reference sensors the reference side of the detector cell is filled with
air and sealed. This eliminates the need to have reference gas constantly
passing through the cell.

NOTE:For instruments equipped with the optional sealed air refer-

ence, there is no REFERENCE inlet or reference VENT port.

It is highly recommended that the same cylinder of gas be used for both

the REFERENCE gas and the ZERO gas.

Pressure, flow, and safety considerations are the same as prescribed for

the SAMPLE gas, above.