beautypg.com

Teledyne 2010A - Split architecture thermal conductivity analyzer User Manual

Page 48

background image

4 Operation

Model 2010A

4-14 Part I

Teledyne Analytical Instruments

####.## % H2 — N2
Slope=#.### S—Zero

Generally, you have a good zero when Slope is less than 0.05 ppm/s for

about 30 seconds.

Once zero settling completes, the information is stored in the analyzer’s

memory, and the instrument automatically returns to the

Analyze

mode.

4.4.1.3 Cell Failure

Cell failure in the 2010A is usually associated with inability to zero the

instrument with a reasonable voltage differential across the Wheatstone bridge. If
this should ever happen, the 2010A system alarm trips, and the VFD displays a
failure message.

Cell cannot be balanced
Check your zero gas

Before replacing the sensor:

a. Check your zero gas to make sure it is within specifications.

b. Check for leaks downstream from the sensor, where contamina-

tion may be leaking into the system.

c. Check flowmeter to ensure that the flow is no more than

200SCCM

d. Check temperature controller board.

e. Check gas temperature.

If none of the above as indicated, the sensor may need to be replaced.

Check warranty, and contact Analytical Instruments Customer Service.

4.4.2 Span Cal

The

Span

button on the front panel is used to span calibrate the analyzer.

Span calibration can be performed in either the automatic or manual mode.

CAUTION: If you are spanning the Cal Range by itself (multiple

application analyzers only), use manual mode
zeroing.

If you want to calibrate ALL of the ranges at once
(multiple application analyzers only), use
auto mode
spanning in the Cal Range.

Make sure the span gas is flowing to the instrument.