2 Installation

Model 2010A

2-2: Part II

Teledyne Analytical Instruments

Figure 3-1: 2010 Analysis Unit with Auto-Cal & Gas Panel options

2-3

Sample System Design

Gas Connector and Selector Panels for specific applications are avail-

able at additional cost . These panels are designed to substitute a standard
front panel.

For those customers wishing to incorporate their own sample system,

electronic input/output ports are provided on the rear panel for the operation
of solenoid valves under the complete control of the Model 2010A electron-
ics. See section 3.3. The recommended system piping schematic is included
among the drawings at the rear of the manual

For best results, use the recommended piping system. Select a

flowmeter that can resolve 0.08 scfh (40-50 cc/min) for the reference path of
the analyzer, and select a flowmeter that can resolve 0.3 scfh (150 cc/min)
for the sample path of the analyzer.