Thermal Conductivity Analyzer

Part I: Control Unit

Part I 4-5

Teledyne Analytical Instruments

4.3.1 Setting the Display

Contrast Function is DISABLED

If you cannot read anything on the VFD after first powering up:

1. Observe LED readout.

a. If LED meter reads 8.8.8.8.8., go to step 3.

b. If LED meter displays anything else, go to step 2.

2. Press I/O button twice to turn Analyzer OFF and ON again. LED

meter should now read 8.8.8.8.8.. Go to step 3.

4.3.2 Setting up an AUTO-CAL

When proper automatic valving is connected (see chapter 3, installation),

the Analyzer can cycle itself through a sequence of steps that automatically zero
and span the instrument.

Note: Before setting up an AUTO-CAL, be sure you understand the

Zero and Span functions as described in section 4.4, and
follow the precautions given there.

Note: If you require highly accurate AUTO-CAL timing, use external

AUTO-CAL control where possible. The internal clock in the
Model 2010A is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.

Note: If all your ranges are for the same gas application, then AUTO-

CAL will calibrate whichever range you are in at the scheduled
time for automatic calibration.

Note: If your ranges are configured for different applications, then

AUTO-CAL will calibrate all of the ranges simultaneously (by
calibrating the Cal Range).

To setup an Auto–Cal cycle:

Choose

System

from the Function buttons. TheVFD will display five

subfunctions.

CONTRAST AUTO—CAL
PWD LOGOUT MORE

Contrast Function is DISABLED

(Refer to Section 1.6)

(Refer to Section 1.6)