beautypg.com

Teledyne 2010A - Split architecture thermal conductivity analyzer User Manual

Page 29

background image

Teledyne Analytical Instruments

Thermal Conductivity Analyzer

Part I: Control Unit

Part I 3-

(Reset by pressing I/o button to remove power. Then

press I/o again and any other button EXCEPT Sys-

tem to resume.

Further detail can be found in chapter 4, section 4-5.

Figure 3-: Types of Relay Contacts

3.3.5 Digital Remote Cal Inputs

Accept 0 V (off) or 24

V dc (on) inputs for remote control of calibra-

tion. (See Remote Calibration Protocol below.)
Zero:

Floating input. 5 to 24 V input across the + and – terminals

puts the analyzer into the Zero mode. Either side may be

grounded at the source of the signal. A synchronous signal

must open and close the external gas control valves appro-

priately. See 3.3.9 Remote Probe Connector. (With the –C

option, the internal valves operate automatically.)

Span:

Floating input. 5 to 24 V input across the + and – terminals

puts the analyzer into the Span mode. Either side may be

grounded at the source of the signal. A synchronous signal

must open and close the external gas control valves appro-

priately. See 3.3.9 Remote Probe Connector. (With the –C

option, the internal valves operate automatically.)

Cal Contact: This relay contact is closed while analyzer is spanning

and/or zeroing. (See Remote Calibration Protocol below.)

Remote Calibration Protocol: To properly time the Digital Remote

Cal Inputs to the Model 2010A Analyzer, the customer's controller must

monitor the Cal Relay Contact.

When the contact is OPEN, the analyzer is analyzing, the Remote Cal

Inputs are being polled, and a zero or span command can be sent.