beautypg.com

Teledyne 2010A - Split architecture thermal conductivity analyzer User Manual

Page 47

background image

Thermal Conductivity Analyzer

Part I: Control Unit

Part I 4-13

Teledyne Analytical Instruments

The beginning zero level is shown in the upper left corner of the display. As

the zero reading settles, the screen displays and updates information on Slope=
in percent/second (unless the Slope starts within the acceptable zero range and
does not need to settle further). The system first does a course zero, shown in
the lower right corner of the screen as C—Zero, for 3 min, and then does a fine
zero, and displays F—Zero, for 3 min.

Then, and whenever Slope is less than 0.01 for at least 3 min, instead of

Slope you will see a countdown: 9 Left, 8 Left, and so fourth. These are
software steps in the zeroing process that the system must complete, AFTER
settling, before it can go back to

Analyze

. Software zero is indicated by S–

Zero in the lower right corner.

####.## % H2 — N2
4 Left=#.### S—Zero

The zeroing process will automatically conclude when the output is within

the acceptable range for a good zero. Then the analyzer automatically returns to
the

Analyze

mode.

4.4.1.2

Manual Mode Zeroing

Press

Zero

to enter the

Zero

function. The screen that appears allows you

to select between automatic or manual zero calibration. Use the

Δ∇ keys to

toggle between AUTO and MAN zero settling. Stop when MANUAL appears,
blinking, on the display.

Select zero
mode: MANUAL

Press

Enter

to begin the zero calibration. After a few seconds the first of

three zeroing screens appears. The number in the upper left hand corner is the
first-stage zero offset. The microprocessor samples the output at a predeter-
mined rate.

####.## % H2 — N2
Zero adj:2048 C—Zero

The analyzer goes through C–Zero, F–Zero, and S–Zero. During C–Zero

and F–Zero, use the

ΔΔΔΔΔ∇ keys to adjust displayed Zero adj: value as close as

possible to zero. Then, press

Enter

.

S–Zero starts. During S–Zero, the Microcontroller takes control as in Auto

Mode Zeroing, above. It calculates the differences between successive sam-
plings and displays the rate of change as Slope= a value in parts per million per
second (ppm/s).