background image

4 Operation

Model 2010A

4-4 Part I

Teledyne Analytical Instruments

4.3

The System Function

The subfunctions of the

System

function are described below. Specific

procedures for their use follow the descriptions:

AUTO-CAL: Used to define an automatic calibration sequence
and/or start an AUTO-CAL.

PWD: Security can be established by choosing a 3 digit
password (PWD) from the standard ASCII character set. Once a
unique password is assigned and activated, the operator MUST
enter the UNIQUE password to gain access to setup functions
which alter the instrument's operation.

LOGOUT: Logging out prevents an unauthorized tampering
with analyzer settings.

MORE: Select and enter MORE to get a new screen with
additional subfunctions listed.

MODEL: Displays Manufacturer, Model, and Software Version
of instrument.

APPLICATION: A restricted function, not generally accessed by
the end user. Used to define up to three analysis ranges and a
calibration range (including impurity, background, low end of range,
high end of range, and % or ppm units).

SELF-TEST: The instrument performs a self-diagnostic test to
check the integrity of the power supply, output boards, sensor
cell, and preamplifiers.

ALGORITHM: A restricted function, not generally accessed by the
end user. Used to linearize the output for the range of interest.