background image

4 Operation

Model 2010A

4-10 Part I

Teledyne Analytical Instruments

RUNNING DIAGNOSTIC
Testing Preamp — Cell

When the testing is complete, the results are displayed.

Power: OK Analog: OK
Cell: 2 Preamp: 3

The module is functioning properly if it is followed by OK. A number

indicates a problem in a specific area of the instrument. Refer to Chapter 5
Maintenance and Troubleshooting
for number-code information. The results
screen alternates for a time with:

Press Any Key
To Continue...

Then the analyzer returns to the initial System screen.

4.3.6 The Model Screen

Move the < > arrow key to MORE and press

Enter

. With MODEL

blinking, press

Enter

. The screen displays the manufacturer, model, and soft-

ware version information.

4.3.7 Checking Linearity with ALGORITHM

From the

System

Function screen, select ALGORITHM, and press

Enter

.

sel rng to set algo:
—> Ш1 Ш2 Ш3 <—

Use the < > keys to select the range: 01, 02, or 03. Then press

Enter

.

Gas Use: H2 — N2
Range: Ø — 10%

Press Enter again.

Algorithm setup:
VERIFY SET UP

Select and

Enter

VERIFY to check whether the linearization has been

accomplished satisfactorily.

Dpt INPUT OUTPUT
Ш Ш.ШШ Ш.ШШ