beautypg.com

Maintenance – Teledyne 2010A - Split architecture thermal conductivity analyzer User Manual

Page 65

background image

Thermal Conductivity Analyzer

Part I: Control Unit

Part I 5-1

Teledyne Analytical Instruments

(Continued)

Maintenance

5.1

Routine Maintenance

Aside from normal cleaning and checking for leaks at the gas connec-

tions, routine maintenance is limited to replacing fuses, and recalibration. For
recalibration, see Section 4.4 Calibration.

WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS

MANUAL.

5.2

System Self Diagnostic Test

1. Press the

System

button to enter the system mode.

2. Use the < > arrow keys to move to More, and press

Enter

.

3. Use the < > arrow keys to move to Self-Test, and press

Enter

.

The following failure codes apply:

Table 5-1: Self Test Failure Codes

Power

0

OK

1

5 V Failure

2

15 V Failure

3

Both Failed

Analog

0

OK

1

DAC A (0–1 V Concentration)

2

DAC B (0–1 V Range ID)

3

Both Failed