beautypg.com

Teledyne 2010A - Split architecture thermal conductivity analyzer User Manual

Page 49

background image

Thermal Conductivity Analyzer

Part I: Control Unit

Part I 4-15

Teledyne Analytical Instruments

4.4.2.1

Auto Mode Spanning

Observe all precautions in sections 4.4 and 4.4.2, above. Press

Span

to

enter the span function. The screen that appears allows you to select whether the
span calibration is to be performed automatically or manually. Use the

ΔΔΔΔΔ∇

arrow keys to toggle between AUTO and MAN span settling. Stop when
AUTO appears, blinking, on the display.

Select span
mode: AUTO

Press

Enter

to move to the next screen.

Span Val: 2Ш.ШШ %
To begin span

Use the < > arrow keys to toggle between the span concentration value

and the units field (%/ppm). Use the

ΔΔΔΔΔ∇ arrow keys change the value and/or the

units, as necessary. When you have set the concentration of the span gas you are
using, press

Enter

to begin the Span calibration.

####.##% H2 — N2
Slope=#.### Span

The beginning span value is shown in the upper left corner of the display.

As the span reading settles, the screen displays and updates information on
Slope. Spanning automatically ends when the span output corresponds, within
tolerance, to the value of the span gas concentration. Then the instrument auto-
matically returns to the analyze mode.

4.4.2.2

Manual Mode Spanning

Press

Span

to start the

Span

function. The screen that appears allows

you to select whether the span calibration is to be performed automatically or
manually.

Select span
mode: MANUAL

Use the

Δ∇ keys to toggle between AUTO and MAN span settling. Stop

when MAN appears, blinking, on the display. Press

Enter

to move to the next

screen.

Span Val: 2Ш.ШШ %
To begin span

Use the < > arrow keys to toggle between the span concentration value

and the units field (%/ppm). Use the

ΔΔΔΔΔ∇ arrow keys change the value and/or the