4 Operation

Model 2010A

4-20 Part I

Teledyne Analytical Instruments

the next lower range, the instrument switches to the lower range. A correspond-
ing shift in the DC concentration output, and in the range ID outputs, will be
noticed.

The autoranging feature can be overridden so that analog output stays on a

fixed range regardless of the contaminant concentration detected. If the concen-
tration exceeds the upper limit of the range, the DC output will saturate at 1 V dc
(20 mA at the current output).

However, the digital readout and the RS-232 output of the concentration

are unaffected by the fixed range. They continue to read beyond the full-scale
setting until amplifier saturation is reached. Below amplifier saturation, the
overrange readings are accurate UNLESS the application uses linearization over
the selected range.

The concentration ranges can be redefined using the

Range

function

Manual screen, and the application gases can be redefined using the

System

function, if they are not already defined as necessary.

CAUTION: Redefining applications or ranges might require

relinearization and/or recalibration.

To setup automatic ranging:

Press

Range

key to start the Range function.

Select range
mode : AUTO

Note: If all three ranges are currently defined for different applica-

tion gases, then the above screen does not display (because
mode must be manual).

If above screen displays, use the

Δ∇ arrow keys to Select AUTO, and

press

Enter.

Press

Escape

to return to the previous Analyze Function.