Thermal Conductivity Analyzer

Part I: Control Unit

Part I 4-19

Teledyne Analytical Instruments

In the AUTO screen, you are further allowed to select which gas application

(PREVIOUSLY defined in

System

function) to run.

4.6.1 Manual (Select/Define Range) Screen

The Manual range-switching mode allows you to select a single, fixed

analysis range. It then allows you to redefine the upper and lower limits, for the
range.

Press

Range

key to start the Range function.

Select range
mode: MANUAL

Note: If all three ranges are currently defined for different applica-

tion gases, then the above screen does not display (because
mode must be manual). Instead, the VFD goes directly to the
following screen.

If above screen displays, use the

Δ∇ arrow keys to Select MANUAL, and

press

Enter.

Select range to run
—> Ш1 Ш2 Ш3 CAL<—

Use the < > keys to select the range: 01, 02, 03, or CAL. Then press

Enter

.

Gas use: H2 — N2
Range: Ø — 10 %

Use the < > keys to toggle between the Range: low-end field and the

Range: high-end field. Use the

ΔΔΔΔΔ∇ keys to change the values of the fields.

Press

Escape

to return to the previous screen to select or define another

range.

Press

Enter

to return the to the

Analyze

function.

4.6.2 Auto (Single Application) Screen

Autoranging will automaticaly set to the application that has at least two

range setup with the same gases.

In the autoranging mode, the microprocessor automatically responds to

concentration changes by switching ranges for optimum readout sensitivity. If the
upper limit of the operating range is reached, the instrument automatically shifts
to the next higher range. If the concentration falls to below 85% of full scale of